This is the company logo Skip to Navigation | Skip to Content
Call Lives Ireland on 1850 555 800
 



   
Technical Specifications:
XpertVal® is a conventional Tc based system for thermal mapping and validation:

Efficient – Tc wires are installed on 3 x 20 channels module with cold junction compensation

Reliable – rugged data logger used in Industrial harsh environment, data displayed real time

Accurate: high accuracy of Tc after pre-calibration (+/-0.2°C)

Data integrity – FDA 21 CFR Part II software

+ Network ready: multiple users may simultaneously share the same data base
  Suitable applications:

Autoclaves
Sterilisation tunnels
Freeze-dryers (lyo cabinets)
Cold Room
ICH stability rooms
Ovens, freezers, refrigerators
Warehouses, incubators
SIP



 
System Type                 60 Channels
Sensors
Sensor Length

Sensor type
Temperature Range

Accuracy
Memory capacity

Tc Tip

Sampling Rate

Software
Module
Close loop calibration


Reference Probes

Dry Blocks



Calibrations

Thermocouples
standard: - 10 m (33 feet)
Customisable: - custom length
all type of thermocouples, most used: T, J, K
- teflon: -200°C to +200°C
- kapton: -200°C to +400°C
+/-0.2°C after pre-calibration
no limitation, data is stored in computer
Backup memory: USB storage device compatible
teflon: teflon tip to prevent moisture on the wire
Kapton: no tip
typical: 10 sec. for 20 channels (high resolution)
Fast scan: 1 sec. for 20 channels (low resolution)
21 CFR Part II
The module accepts several inputs: 4 – 20 mA / 0-10 V / RTD (PT-100 2, 3, or 4 wires)
always needed when using thermocouples
system use temperature block and reference probe, software controlled
reports generated automatically by the software
IRTD
Hart 1502
Lives: PTC 125 : -90°C to 125°C / RTC 157 : -45°C to 155°C / RTC 159 : -100°C to 155°C
RTC 700 : +23°C to 700°C
Fluke 9190: -90 to 140°C; Fluke 9170: -45 to 140°C; Fluke 9172: 35 to 425°C
Factory: NIST / COFRAC
User Calibration 



   




















XpertLog® simplifies and makes the thermal validation process more efficient:

High efficiency: Easy to program and install, pre-calibration not needed

High reliability: no data loss during validation process

High accuracy: ±0.1°C from -90°C to 140°C

Data Integrity: FDA 21 CFR Part II

+ Network ready: multiple users may simultaneously share the same data base
+ Remote factory calibration on users site
  Suitable applications:

Autoclaves / SIP
Sterilisation tunnels
Freeze-dryers (lyo cabinets)
Cold Room / ICH stability rooms
Ovens / Freezers /
Refrigerators /
Warehouses / Incubators

Technical Specifications:
Logger Type Temperature                 
Humidity                      
Pressure                     
Sensor - type

-flexible
- rigid
-customised
Range
Accuracy


Size - diameter
-height
Memory Capacity

Battery
Sampling Rate
Battery Lifetime
Internal Clock Drift
4 wires PT-100

PTFE/SS 19" (480mm)
1"(25mm) / 6"(150mm)
up to 39" (1000 mm)
-90°C to 140°C
±0.1°C from -90°C to 140°C


40 mm
30 mm
32,000 data points

lithium batteries (user replaceable)
1 sec to 24 hours
Up to 12 month - based on user usage
4 sec / 24 h @ 23°C
4 wires PT-100
capacitive

1" (25mm)
19" (480mm)
-35°C to 100°C; 10 - 95 % RH
±0.1°C from -35°C to 100°C;
±1.2 % RH @ 23°C;
±2.0 % RH over range 20 to 45°C*
40mm
70mm
16,000rh / 16,000 temp





4 wires PT-100
absolute pressure

1" (25mm)
19" (480mm)
-40°C to 140°C; 0.01 to 5 bar abs
±0.1°C from -40°C to 140°C;
±10 mbar @ 0.01; 1&2 bar abs
±15 mbar @ 0.1; 0.2 & 3 bar abs
40mm
100mm
16,000p / 16,000 temp





Docking Station 

Material
Enclosure
Calibrations

Certification 




USB to computer: Available models: 4, 8, 12 and 20 ports
Program and read unlimited numbers of loggers, regardless of the number of ports
Ketron peek - FDA compliant
IP-68
Factory: traceable NIST / COFRAC - ITS-90 coefficients resides in the internal memory
User: Close loop user calibration, on user site, using provided validation software XpertLog®
EMC : EN 60601-1-2/2007
RF: EN 300440-2 v1.3.1 (2009-03)
Health Safety : EN 60950-1/2006
* Temperature compensated between 20 to 45°C - can be extended to higher temperatures



   



XpertLog® GDP

covers all the applications between
-40°C to 85°C

offers the best accuracy in its class:
±0.1°C

user friendly

data integrity: FDA 21 CFR Part II compliant

+ Network ready: multiple users may simultaneously share the same data base
  Suitable applications:

Refrigerators 

Cold Room
Warehouses

Technical Specifications:
Logger Type Temperature         
Humidity   
Sensor Type

- flexible
- rigid
- customised
Range
Accuracy


Weight
Battery
Diameter
Height
Sampling Rate
Memory Capacity
Battery Lifetime
Internal Clock drift
Docking station

Material
Enclosure
Calibrations                          

4 Wires PT-100

PTFE 19" (480mm)

1" (25mm) / 6" (150mm)
-40°C to 85°C
±0.1°C


20 g
lithium, user replaceable
40mm
27mm
1 sec to 24 hours
32,000
up to 10 month - based on user usage
4 sec / 24 hours @ 23°C
USB to computer
Available models: 4, 8, 12 and 20 ports
FDA compliant
IP67
Factory: traceable NIST / COFRAC - ITS-90
coefficients resides in the internal memory
4 wires PT-100
capacitive

1" (25mm)
PTFE 19" (480mm)
15% to 95% RH
±0.1°C
±2.0% RH @ 23°C (15-95% RH)

50g
lithium, user replaceable
40mm
35mm
1 sec to 24 hours
16,000rh / 16,000 temp









   
















XpertLog® RF elevates thermal validation process to a whole new level:

Highly efficient: increase production efficiency by cutting down validation time

Highly reliable: real time data during validation process

Impressive accuracy: ±0.1°C from -90°C to 140°C

Data Integrity: FDA 21 CFR Part II

+ Network ready: multiple users may simultaneously share the same data base
+ Remote factory calibration on users site

  Suitable applications:

Autoclaves / SIP
Sterilisation tunnels
Freeze-dryers (lyo cabinets)
Cold Room / ICH stability rooms
Ovens / Freezers /
Refrigerators /
Warehouses / Incubators, etc



 
Technical Specifications:
Logger Type
   Temperature                 
    Humidity                      
  Pressure                     

Sensor - type

-flexible
- rigid
-customised
Range
Accuracy


Wireless range
Weight
Antenna
Battery
Size - diameter
-height
Sampling Rate
Logger Update Rate   
Memory Capacity

4 wires PT-100

PTFE/SS 19" (480mm)
1"(25mm) / 6"(150mm)
up to 39" (1000 mm)
-90°C to 140°C
±0.1°C from -90°C to 140°C


400 feet (121m) line of sight
85g battery included
internal
2 lithium batteries, user replaceable)
60mm
30mm
1 sec to 24 hours
5 sec to 24 hours - user configurable
32,000 data points
4 wires PT-100
capacitive

1" (25mm)
19" (480mm)
-35°C to 100°C; 10 - 95 % RH
±0.1°C from -35°C to 100°C;
±1.2 % RH @ 23°C;
±2.0 % RH over range 20 to 45°C*
400 feet (121m) line of sight
105g battery included
internal
2 lithium batteries, user replaceable)
60mm
76mm
1 sec to 24 hours
5 sec to 24 hours - user configurable
16,000p / 16,000 temp
4 wires PT-100
absolute pressure

1" (25mm)
19" (480mm)
-40°C to 140°C; 0.01 to 5 bar abs
±0.1°C from -40°C to 140°C;
±10 mbar @ 0.01; 1&2 bar abs
±15 mbar @ 0.1; 0.2 & 3 bar abs
400 feet (121m) line of sight
155g battery included
internal
2 lithium batteries, user replaceable)
60mm
85mm
1 sec to 24 hours
5 sec to 24 hours - user configurable
16,000p / 16,000 temp
Battery Lifetime
Internal Clock drift
Access Point
Material
Radio frequency
Enclosure
Calibrations
 




Up to 6 month - based on user usage
4 sec / 24 h @ 23
°C
USB to computer:
Ketron peek - FDA compliant
2.4 Ghz ISM band
IP-68
Factory: traceable NIST / COFRAC - ITS-90 coefficients resides in the internal memory
User: Close loop user calibration, on user site, using provided validation software XpertLog®
EMC : EN 60601-1-2/2007
RF: EN 300440-2 v1.3.1 (2009-03)
Health Safety : EN 60950-1/2006
* Temperature compensated between 20 to 45°C - can be extended to higher temperatures



 
Technical Specifications:
Features:

- create reliable reports - all calculations are performed after the data is downloaded from the loggers

Split the raw data stored inside loggers during several tests in as many tests performed

Create:
  • raw data reports and F value reports
  • trend reports
  • statistics reports
  • image reports
  • steam saturation reports
  • calibration certificate report
  • 3 D reports
Assign certifications at the performed tests

Create custom calculations using the "Virtual Channels"

Add comments in the header of the reports
 
 - USB Communication with  - 4 access levels  - close loop calibration
 - data loggers
 - temperature blocks
 - reference probes
 - administrator
 - user
 - reporter
 - reviewer
 - calibration / verification of the sensor before running a test
 - verification of the loggers after running a test
 - fully automated calibration - the calibration equipment
   controlled by the software
- use of templates to create final reports    - generate final reports
 
 - create a library of users / organisations
 - create acceptance criteria templates
 - create a library of equipments




 - all the reports generated by the software merged
   in one single report
 - paginated
 - conclusion of the report on the first page
 - acceptance criteria
 - summary of the results

 - 3D view
 - trends
 - stream saturation
 - images
 - raw data
 - calibration / verification report
 - calibration certificates

       
 XpertLog GDP
 XpertLog Standard
  XpertLog Real Time
  XpertVal
       
XpertLog Software
    Sample Fridge Study